Thermocirculator and Recirculating Heater / Chillers
E3500 Thermocirculator
The E3500 Thermocirculator is a low cost, portable water circulating system for supplying a constant temperature supply for closed and open loop applications at near ambient to +60°C.
Features
- Precise control of heating
- Compact
- Robust and reliable
- No running water to waste
- Ideal for use with the E3100 Critical Point Dryer
The E3500 is suitable for open and closed loop applications and has all controls and liquid connections on the front panel. The temperature controller shows not only the "set" temperature (between ambient and +65°C), but also the actual liquid temperature. The fluid is circulated by a glandless magnetic pump, through an aluminum reservoir.
A cooling coil is mounted in the liquid reservoir. This coil can be directly connected to the main water supply and is primarily used when temperature control near to ambient is required. The electronic control system features a zero voltage switching unit which minimizes mains electrical interference, enabling the E3500 to be used and sited with other sensitive electronic instruments.
Using the E3500 with critical point dryers
The E3500 will give controlled heating of E3100 Critical Point Dryer. It is connected directly to the inlet and outlet of the E3100 water jacket. The temperature of the circulating fluid is pre-set,typically set to 37°C (just above the critical temperature of carbon dioxide).
Barley Leaf Trichomes and stomatal pores on the epidermal surface of a barley (Hordeum vulgare) leaf. Some very fine wax crystallites are also just visible on the surface of the leaf. |
Stomatal Pore on Xerophyte Leaf Surface Critical point dried epidermis of a xerophyte (cactus), demonstrating raised stomatal pores. |
Mature Spruce Wood Critical point dried block of mature spruce wood block, demonstrating transverse, tangential and radial views of tracheids and vessels. |
Specifications
Working temperature range | Ambient to +65°C |
Heating rating | 500W |
Reservoir capacity | 2.2L / 3.5 Pints |
Pump type | Glandless magnetic drive |
Pump motor | Shaded pole |
Pump flow | 270L/h (zero head) / 60 Gal/hr (zero head) |
Pump pressure | 0.25 kg / cm2 / 3.5 psi |
Max pump head (unrestricted) | 2.5 m / 7.5 ft |
Size (unpacked) | 330 x 280 x 150 mm / 13 x 11 x 6" |
Weight | 8 kg / 17.6 lbs |
Electrical | 220-240V/50Hz, 115V/60Hz |
Temperature range | -20°C to +70°C |
Ordering Information
91095 | E3500 Thermocirculator | each | POR | Quote |
E4800 Recirculating Heater/Chillers
Recommended for open and closed loop applications, offering simplicity, reliability and quiet operation. The range includes the E4860, 4870, 4880 and 4890.
Features
- Precise temperature control
- Quiet, efficient operation
- Proven reliability
- Low maintenance
- Environmentally friendly - avoids running water to waste
Some typical applications
- Vacuum coating equipment
- Critical point dryers including the E3100
- Electron microscopes
- Chromatography equipment
- Electrophoresis baths
- Environmental chambers
- Crystal growth apparatus
- Fermentation equipment
- Interferometers
- Photographic baths
- X-ray equipment
- Polarimeters, refractometers
...and many others
Temperature control
Many instruments measuring physical properties depend on accurate control of temperature and in some processes optimum temperature is essential. With the E4800 series, over-cooling (which affects efficiency) is prevented and the water temperature can be accurately controlled over the range -10 to +60°C.
A commonly misunderstood feature of refrigerated systems is in applications where the control temperature is other than at or near room temperature. When the instruments are to be operated at controlled temperatures below ambient, the extraction deteriorates significantly and, as a guide, the compounded change is 4% per degree Celsius. In practice, the refrigerant gas pressure has to be adjusted to optimize the performance at any particular temperature. However, the E4800 series incorporate automatic adjustment valves in the systems.
The E4800 series are "closed loop" and efficiencies are dramatically improved compared with open bath models. They are simple to set up and to operate, and essentially maintenance free.
Choosing the correct heater/chiller
In order to optimize performance from a heater/chiller system, the correct specification must be selected for a particular application. To cool or heat any instrument or system it is important to obtain the following information from the manufacturer:
- Heat load to be dissipated to water, for an electron microscope: diffusion pump heater, lenses, etc
- Flow rate and size of tubing
- Minimum pressure
With this information, consult the Specifications below and select the appropriate heater/chiller. The basic heat load calculation formula is as follows:
Flow rate x weight of fluid x specific heat x D T = Heat Extraction.
Specifications
Heat Extraction Rates (in Watts) | - 20°C | - 10°C | 0°C | +10°C | + 20°C |
E4860 | 75W | 105W | 180W | 300W | 420W |
E4870 | 125W | 250W | 500W | 900W | 1.2kW |
E4880 | 200W | 425W | 700W | 1.6kW | 2.2kW |
E4890 | 350W | 600W | 1.2kW | 2kW | 3kW |
4.5kW Recirculator | 700W | 1kW | 2kW | 3kW | 4.5kW |
6kW Recirculator | 800W | 1.3kW | 2.6kW | 4.5kW | 6kW |
Model | E4860 | E4870 | E4880 | E4890 |
Extraction rate at 20°C | 400W | 1.4kW | 2.2kW | 3kW |
Temperature range | -20°C to +70°C | -20°C to +70°C | -20°C to +70°C | -20°C to +70°C |
Refrigeration (HP) | 1/5 | 1/2 | 3/4 | 1 |
Heater rating | 1kW | 1.5/2.0kW | 2.5kW | 2.5kW |
Max pump flow | 450L/h 200Gal/hr |
450L/h 200Gal/hr |
900L/h 275Gal/hr |
900L/h 275Gal/hr |
Tank capacity | 1.2L | 1.7L | 2.3L | 3.0L |
Max pump pressure psi/bar | 12/60psi 0.7/3bar |
12/60psi 0.7/3bar |
60psi 1.5/3bar |
60psi 1.5/3bar |
Height | 37cm | 45cm | 50cm | 50cm |
Width | 32cm | 38cm | 45cm | 45cm |
Depth | 46cm | 61cm | 62cm | 62cm |
Weight | 40kg | 62kg | 70kg | 82kg |
Water connections | 16mm hose or 1/8 BSP | 16mm hose or 1/8 BSP | 16mm hose or 1/8 BSP | 16mm hose or 1/8 BSP |
Temperature sensor | R.T. Probe | R.T. Probe | R.T. Probe | R.T. Probe |
Optional Attachments
- High pressure pump for E4860 and E4870 (standard in E4880 and E4890)
- Water failure alarm
- Over and under temperature cut out
NOTE: Larger capacity heater/chillers (6kW and 12kW) are available on request - please contact us for further information - Custom-made heater/chiller units
E4860 (91098) | 1/5 HP Recirculating Heater/Chiller | each | POR | Quote |
E4870 (91099) | 1/2 HP Recirculating Heater/Chiller | each | POR | Quote |
E4880 (91090) | 3/4 HP Recirculating Heater/Chiller | each | POR | Quote |
E4890 (91095) | 1 HP Recirculating Heater/Chiller | each | POR | Quote |